High Intensity Magnetic Separator

High intensity magnetic separator is suitable for the laboratory of mineral processing, metallurgy, geology to separate of weak magnetic or strong magnetic minerals with dry method. The maximum magnetic field intensity is up to 14000 Gs

High intensity magnetic separator is suitable for the laboratory of mineral processing, metallurgy, geology to separate of weak magnetic or strong magnetic minerals with dry method.

Working Principle and structure:

The machine adopts double working face with closed magnetic circuit, which is composed of excitation system, transmission mechanical feeding system, and product cleaning and collection system.
Electromagnetic vibration ore feeder will give minerals into the seperation zone, by controlling the amount of excitation current, so that minerals through the magnetic gap automatically divided into magnetic and non-magnetic materials, discharging from the machine, to achieve the sorting effect.

structure of High Intensity Magnetic Separator

01. Rotating magnetic roller (fixed magnetic roller)   02. Moving roller motor   03 control panel
04. Heavy hammer cleaning     06. Gap adjusting bolt        07 ore discharging chute

Main Technical Specifications:
1. Roll size: diameter φ120mm
Width (double working face) 30+30mm
2. Maximum magnetic field intensity: 14000 Gs
In general use, it can be adjusted in the range of 600-14000 Gs (associated with material characteristics).
3. Magnetic pole spacing adjustment range: 4-8mm
4. Size range of test materials :When(magnetic pole spacing) 0-5 mm , Generally applicable particle size: ~ 1.0-3.0mm.
5. Magnetic field adjustment range (for reference, related to material characteristics) :
 When the magnetic pole spacing is 4mm
Strong magnetic field position 5000-14000 Gs
Weak magnetic field position 1600-5000 Gs
     When the spacing is 6mm
Strong magnetic field position 4000-12000 Gs
Weak magnetic field position 1000-4000 Gs
   When the spacing is 8mm
Strong magnetic field position 3000-10000 Gs
Weak magnetic field position 600-3000 Gs
6 reference capacity: 5-25 kg/h
7. Roll speed: 55 r/min
8. Electric feeder: frequency adjustable.
9. Power supply: 220V50Hz The equipment body is well grounded.
10. Power consumption: less than 500VA.
11. Dimensions: length × width × height (360×450×460 mm)
12. Reference weight: ~ 75 kg

 

Relationship curve between magnetic field intensity and current

Relationship curve between magnetic field intensity and current